Hand-held electron spin resonance scanner for subcutaneous oximetry using OxyChip

CONCLUSION: The compact ESR scanner can report pO2 data in ATE phantom from an external position relative to the scanner. Implementing this scanner in preclinical and clinical applications for subcutaneous pO2 measurements is a feasible next phase for this development. This innovative design also has the potential to operate in conjunction with artificial skin graft for wound healing, combining therapeutic and pO2 diagnostic features.PMID:38411265 | DOI:10.1002/mrm.30066
Source: Magnetic Resonance in Medicine - Category: Radiology Authors: Source Type: research