The assessment of the clinical impact of using a single set of radiotherapy planning data for two kilovoltage therapy units

This study reviewed kilovoltage dosimetry and treatment planning scenarios to evaluate the feasibility of using ACPSEM annual QA tolerances to determine whether two units (of the same make and model) were dosimetrically matched. The dosimetric characteristics, such as measured half value layer (HVL), percentage depth dose (PDD), applicator factor and output variation with stand-off distance for each kV unit were compared to assess the agreement. Independent planning data based on the measured HVL for each beam energy from each kV unit was prepared. Monitor unit (MU) calculations were performed using both sets of planning data for approximately 200 clinical scenarios and compared with an overall agreement between units of  <  2%. Additionally, a dosimetry measurement comparison was completed at each site for a subset of nine scenarios. All machine characterisation measurements were within the ACPSEM Annual QA tolerances, and dosimetric testing was within 2.5%. This work demonstrates that using a single set of planning data for two kilovoltage units is feasible, resulting in a clinical impact within published uncertainty.