Feasibility and efficacy of a novel size adjustable cryoballoon for ablation of atrial fibrillation

AbstractThe aim of the present case series was to characterize the feasibility of a novel size adjustable cryoballoon system (PolarX Fit, Boston Scientific, Marlborough, MA, USA). This cryoballoon catheter can be inflated to two different diameters (28 mm and 31 mm) within the same procedure allowing vein adapted PVI. In summary, the novel size adjustable cryoballoon shows similar characteristics as the established versions. The intraprocedural flexibility of balloon size led to employment of the larger variant in the majority of freeze applications. Of note, in all but one procedure, both sizes were employed to ensure optimal occlusion for all veins. This initial series suggests that the size adjustable balloon offers more flexibility of obtain optimal occlusions in particular, in challenging anatomies, including common pulmonary vein ostia.
Source: Journal of Interventional Cardiac Electrophysiology - Category: Cardiology Source Type: research