Physical wedge as a tool for radiochromic film calibration

In this study the feasibility of using dose gradients produced by a physical wedge (PW) for RCF calibration was investigated. The aim was to establish an efficient and reproducible method for calibrating RCF using a PW. Film strips were used to capture the wedge dose profile for five different exposures and the acquired scans were processed to generate corresponding net optical density wedge profiles. The proposed method was compared to the benchmark calibration, following the guidelines for precise calibration using uniform dose fields. The results of the benchmark comparison presented in this paper showed that using a single film strip for measuring wedge dose profile is sufficient for estimating a reliable calibration curve within the recorded dose range. Furthermore, the PW calibration can be extrapolated or extended by using multiple gradients for the optimal coverage of the desired calibration dose range. The method outlined in this paper can be readily replicated using the equipment and expertise commonly found in a radiotherapy center. Once the dose profile and central axis attenuation coefficient of the PW are determined, they can serve as a reference for a variety of calibrations using different types and batches of film. This investigation demonstrated that the calibration curves obtained with the presented PW calibration method are within the bounds of the measurement uncertainty evaluated for the conventional uniform dose field calibration method.PMID:37393128 | ...
Source: Zeitschrift fur Medizinische Physik - Category: Radiology Authors: Source Type: research
More News: PET Scan | Radiology | Study