Evaluation, Validation, and Implementation of the Idylla System as Rapid Molecular Testing for Precision Medicine
We describe our evaluation, validation, and implementation of this system for routine testing of BRAF, EGFR, KRAS, and NRAS using formalin-fixed, paraffin-embedded cancer samples. All four Idylla test systems showed excellent concordance with reference methods. The analytical sensitivity ranged from 94.66% to 100%, depending on the cartridge, and specificity was 100%.
Source: Journal of Molecular Diagnostics - Category: Pathology Authors: Huiya Huang, Stephanie Springborn, Kiefer Haug, Kaitlyn Bartow, Hasan Samra, Smitha Menon, Alexander C. Mackinnon Tags: Regular article Source Type: research