Steam sterilization effect on the accuracy of friction-style mechanical torque limiting devices

Conclusion: Steam sterilization effect differs between target torque and measured peak values with an increase trend. The peak torque values showed a significant decrease for BioHorizons, while a significant increase was noted for Astra Tech and no significant change in Dr. Idhe group after sterilization. Clinical Implication: Within the limitation of this study the torque output of each individual device deviated in varying degrees from target torque values. However, the majority of the new frictional-style devices tested in this study, delivered fairly consistent torque output within 10% of their preset target values after sterilization. Astra Tech devices were the only one showing more than 10% difference from their torque values in 5% of the measurements. Combined effects of sterilization and aging still needs to be determined.
Source: Table of Contents : Indian Journal of Dental Research : 2006 - 17(3) - Category: Dentistry Authors: Source Type: research
More News: Dentistry | Study