Use of an Image Acquisition Stabilizer Improves Sidestream Dark Field Imaging of the Serosa during Open Gastrointestinal Surgery

Aim: To investigate whether an image acquisition stabilizer (IAS) mounted on the sidestream dark field camera (SDF) during gastrointestinal surgery improves image stability and acquisition.Methods: Serosal SDF imaging was compared with SDF imaging combined with an IAS (SDF + IAS) during gastrointestinal surgery. Stability was assessed as the image drift in pixels and the time to obtain stable images. The success rate was determined as the percentage of analyzable images after recording. The effect of negative pressure from the IAS was determined during single-spot measurements and by comparing microvascular parameters between groups. Data are presented as mean ± SD.Results: Sixty serosal measurements were performed per group; 87% were successful in the SDF group and 100% in the SDF + IAS group (p = 0.003). Image drift in the SDF group was 148 ± 36 versus 55 ± 15 pixels in the SDF + IAS group; p
Source: Journal of Vascular Research - Category: Research Source Type: research