A Direct Current-Sensing VCO-Based 2nd-Order Continuous-Time Sigma-Delta Modulator for Biosensor Readout Applications

A second-order voltage-controlled oscillator (VCO)-based continuous-time sigma-delta modulator (CTSDM) for current-sensing readout applications is proposed. Current signals from the sensor can directly be quantized by the proposed VCO-based CTSDM, which does not require any extra trans-impedance amplifiers. With the proportional-integral (PI) structure and a VCO phase integrator, the capability of second-order noise shaping is available to reduce the in-band quantization noise. The PI structure can be simply realized by a resistor in series with the integrating capacitor, which can reduce the architecture complexity and maintain the stability of the system. The current-steering digital-to-analog converter with tail and sink current sources is used on the feedback path for the subtraction of the current-type input signal. All the components of the circuit are scaling friendly and applicable to current-sensing readout applications in the Internet of Things (IoT). The proposed VCO-based CTSDM implemented in a 0.18-μm standard CMOS process has a measured signal-to-noise and distortion ratio (SNDR) of 74.6 dB at 10 kHz bandwidth and consumes 44.8 μw only under a supply voltage of 1.2 V, which can achieve a Figure-of-Merit (FoM) of 160.76 dB.
Source: IEEE Transactions on Biomedical Circuits and Systems - Category: Biomedical Engineering Source Type: research