Novel multileaf collimator designs with tongues and grooves in leaf end and Monte Carlo simulations

In this study, we proposed 2 new multileaf collimator leaf designs to eliminate leaf gaps for closed leaf pairs so that radiation leakage can be avoided. In the new designs, multi tongues and grooves were added to the conventional multileaf collimators leaf ends. Thus, when a pair of leaves closed, tongues of a leaf can enter grooves of its opposing leaf. Consequently, there would be no radiation leakage through closed leaves. One design was named finger-shaped MLC, and another design with doubled leaf end thickness was named hand-shaped MLC.
Source: Medical Dosimetry - Category: Radiology Authors: Tags: Research Article Source Type: research