Evaluation of the RadCalc collapsed cone dose calculation algorithm against measured data

This work describes the experimental validation of the RadCalc (Lifeline software Inc, Tyler) collapsed cone dose calculation algorithm against measured data for a range of scenarios. 6 MV photon beam data were measured in a large water tank on a Varian TrueBeam linear accelerator. These were input into the RadCalc software, in conjunction with head geometry and output calibration information, then used to create a collapsed cone beam model. The model performance was assessed by comparison against measurement, using a selection of homogeneous and inhomogeneous geometries not incorporated into the original beam model.
Source: Medical Dosimetry - Category: Radiology Authors: Tags: Medical Physics Contribution Source Type: research