Dose verification of virtual bolus application for helical tomotherapy
The objective of the study is to verify the dose delivered on helical tomotherapy based on treatment plan with varying virtual bolus (VB) thickness. The target was localized on the ArcCHECK image by 3 mm margin from the phantom surface. The dimension of target, which includes the ArcCHECK's detectors, with the 4.0 cm width and length 12.0 cm along the phantom The 5 treatment plans were generated, 1 plan without VB application (NoVB) and the 4 plans with varying of VB thickness on the phantom surface by 0.5 cm (VB0.5), 1.0 cm (VB1.0), 1.5 cm (VB1.5), and 2.0 cm (VB2.0), in treatment planning but absent during irradiation.
Source: Medical Dosimetry - Category: Radiology Authors: Saw Yu Nwe, Imjai Chitapanarux, Wannapha Nobnop Source Type: research