Esophageal Perforation after Using a Single-Use Disposable Duodenoscope

We describe a case of esophageal perforation after using a single-use disposable duodenoscope (EXALT Model D; Boston Scientific Corporation, Marlborough, MA, USA). To our knowledge, this is the first reported case of an esophageal perforation since FDA approval of this device in December 2019. ERCP was performed with the EXALT Model D single-use duodenoscope (Boston Scientific Corporation) by an experienced gastroenterologist. During the procedure, gentle force applied through the gastroesophageal junction caused a liner perforation in the distal esophagus. An esophageal stent was placed with satisfactory wound healing and no fistula formation. There have been a few reports in the last 2 years showing promising results using this device; however, the differences in the tactile feedback, navigation, and pushability of the device may make it prone to unintended consequences.Case Rep Gastroenterol 2021;15:972 –977
Source: Case Reports in Gastroenterology - Category: Gastroenterology Source Type: research