Safety analysis of intermittent hemodialysis in patients with continuous flow left ventricular assist devices

In this study, the effect of HD was observed in such patients. Between June 2009 and October 2012, 139 patients received LVADs, of which 10 patients (7%) required intermittent HD postoperatively. The mean age of the patients was 53 ± 14 years and 90% were men. A total of 281 dialysis sessions were administered amounting to 1025 hours of dialysis. The mean systolic blood pressure monitored with Doppler device was 97 ± 18 mmHg. Dialysis durations averaged 218 ± 18 minutes. Mean blood flow rate was 334 ± 38 cc/min, and 2.6 ± 1.1 L was ultrafiltrated during each session. Only 15 (5.3%) sessions were interrupted or terminated in six patients. The reasons for termination were symptomatic hypotension—6 (2.1%), asymptomatic hypotension—3 (1%), ventricular tachycardia—1 (0.36%), dialysis machine malfunction—2 (0.7%), low phosphorus—2 (0.7%), and abdominal cramps—1 (0.36%). Volume expansion was necessary on three occasions. Low‐flow device alarms were registered during two (0.71%) sessions. The results showed no serious adverse effects or deaths.
Source: Hemodialysis International - Category: Hematology Authors: Tags: Case Report Source Type: research
More News: