SunMed Holdings, LLC Issues a Nationwide Recall of STAT-Check and Medline Manual Resuscitator Bags

On May 29, 2018, SunMed Holdings, LLC initiated a nationwide recall of 18,808 units of STAT-Check and Medline resuscitator bags which were distributed between February 1, 2018 to May 13, 2018. The patient port retaining ring of the affected bags may not fully seat which may allow the patient port to detach during use. If this takes place, the resuscitator bags may not deliver air to the patient and result in a delay in treatment and life-threatening health consequences. There have been no reported injuries to date.
Source: Food and Drug Administration - Category: Food Science Source Type: news