Defining an equivalent geometry for Monte Carlo calculations of a high-purity Ge detector with high relative efficiency using a genetic algorithm

Appl Radiat Isot. 2022 May 18;186:110295. doi: 10.1016/j.apradiso.2022.110295. Online ahead of print.ABSTRACTDetailed geometric information of a high-purity Ge (HPGe) detector is a very important issue for Monte Carlo simulation of the detector. Commonly, users have no geometric information about the detector and information given by the manufacturer is not completely valid for simulation. An equivalent geometry of detector, the parameters of which can be used for Monte Carlo simulation, is optimised using a genetic algorithm for a large-volume HPGe detector in this study. A mixed-point gamma calibration standard, emitting 12 useful gamma-radiation energies within 59.5-1836.1 keV, is placed at 74 different locations around the detector for this purpose. A high-quality solution is generated starting from an initial population of randomly-generated detector geometries using a genetic algorithm. Fitness of each geometry is obtained by comparing full energy peak efficiencies computed by Monte Carlo simulation with experimental values for each energy and position. Efficiencies with relative errors less than 5% for high energies and less than 7% for lower energies, except 59.5 keV, are obtained using optimised equivalent geometry parameters for the Monte Carlo simulation. Also, the necessity of using crystal dimensions smaller than real dimensions for Monte Carlo simulations of high-volume HPGe detectors is discussed. In addition, for Monte Carlo simulation of high-volume HPGe dete...