Virtual Unenhanced Images: Qualitative and Quantitative Comparison Between Different Dual-Energy CT Scanners in a Patient and Phantom Study

Objectives Materials and Methods Forty-four patients with clinical contrast-enhanced abdominal examinations on each of the 3 DECT scanner types and a phantom scanned with the same protocols were included in this retrospective study. Qualitative and quantitative assessment was performed on VUE images. Quantitative evaluation included measurement of attenuation and image noise for various tissues and the phantom. Virtual unenhanced image attenuation and noise were compared between scanner types, and intrapatient interscanner reproducibility of virtual unenhanced image attenuation was calculated as the percentage of measurement pairs with an interscanner difference ≤ 10 HU. Image quality, noise, sharpness, and iodine elimination were assessed qualitatively by 2 radiologists. Results Significant interscanner differences in VUE attenuation and noise were found in all tissues. dlDECT and rsDECT showed significantly higher VUE attenuation than dsDECT in the aorta, portal vein, and kidneys (P