Comparison of the TOFscan and the TOF-Watch SX during pediatric neuromuscular function recovery: a prospective observational study

ConclusionsTOFscan reliably demonstrated lack of bias and good concordance with TOF-Watch SX throughout the neuromuscular recovery, especially when normalized. Despite technical limitations, the two devices were unbiased along the path of spontaneous and pharmacological reversal in pediatric patients.Trial registrationClinicalTrials.gov NCT03775603. Registered on 13 March 2018
Source: Perioperative Medicine - Category: Surgery Source Type: research