Noninvasive Beat-To-Beat Stroke Volume Measurements to Determine Preload Responsiveness During Mini-Fluid Challenge in a Swine Model: A Preliminary Study

The objectives of this paper are to compare EIT with an invasive pulse contour analysis (PCA) method in measuring SV during mini-fluid challenge in animals and determine preload responsiveness with EIT. Five pigs were anesthetized and mechanically ventilated. After removing 25% to 30% of the total blood from each animal, multiple fluid injections were conducted. The EIT device successfully tracked changes in SV beat-to-beat during varying volume states, i.e., from hypovolemia and preload responsiveness to target volume and volume overload. From a total of 50 100-mL fluid injections on five pigs (10 injections per pig), the preload responsiveness value was as large as 32.3% in the preload responsiveness state while in the volume overload state it was as low as −4.9%. The bias of the measured SV data using EIT and PCA was 0 mL, and the limits of agreement were ±3.6 mL in the range of 17.6 mL to 51.0 mL. The results of the animal experiments suggested that EIT is capable of measuring beat-to-beat SV changes during mini-fluid challenge and determine preload responsiveness. Further animal and clinical studies will be needed to demonstrate the feasibility of the EIT method as a new tool for fluid management.
Source: Shock - Category: Emergency Medicine Tags: Basic Science Aspects Source Type: research