Experimental evaluation of the injection force exerted in intraocular lens delivery with syringe-type injectors

This study aims to obtain the main parameters that affect the injection force exerted in the ejection of an intraocular lens (IOL) through syringe-type injectors. For that purpose, ejection tests were carried out in vitro, measuring the resistance force throughout the entire delivery process. The effect of IOL material, haptic design, IOL thickest area and ophthalmic viscosurgical device (OVD) was studied by ejecting seven IOLs with four syringe-type injectors of different sizes, 3.0, 2.2 and 1.8 mm. In all injectors, plate hydrophilic IOLs present the lowest resistance forces; hydrated C-loop hydrophobic IOLs present higher forces and the C-loop hydrophobic IOL in dry conditions presents the highest resistance forces. All IOLs could be properly delivered with an injector size of 2.2 mm, making injector sizes of 3.0 mm outdated. The injector size of 1.8 mm damaged several IOLs. IOL material and cartridge nozzle size were the most influential parameters in IOL delivery. IOL thickest area was also relevant but in a lesser extent whereas IOL haptic design was not as relevant.PMID:34492403 | DOI:10.1016/j.jmbbm.2021.104793