Photon-Counting Multienergy Computed Tomography With Spectrally Optimized Contrast Media for Plaque Removal and Stenosis Assessment

Purpose The aim of this study was to systematically evaluate the potential to combine investigational contrast media with spectrally optimized energy-thresholding of photon-counting detector computed tomography (PCCT) for subtraction of calcified plaques in a coronary artery stenosis phantom. Methods A small vessel phantom containing 3 fillable tubes (diameter, 3 mm each) with calcified plaques was placed into an anthropomorphic chest phantom. The plaques had incremental thicknesses ranging from 0.3 to 2.7 mm, simulating vessel stenoses ranging from 10% to 90% of the lumen diameter. The phantom was filled with 5 different investigational contrast media (iodine, bismuth, hafnium, holmium, and tungsten) at equal mass concentrations (15 mg/mL) and was imaged on a prototype PCCT at 140 kVp using optimized, contrast media–dependent energy thresholds. Contrast maps (CMs) were reconstructed for each contrast medium by applying a linear 2-material decomposition algorithm. Image noise magnitude and noise texture of CM were compared among the contrast media using the noise power spectrum. Two blinded readers independently rated the vessel lumen visualization on short-axis and the overall subjective image quality on long-axis CM relative to iodine as the reference standard. Four readers determined the highest degree of stenosis that could be assessed with high diagnostic confidence on long-axis CM. Results Average image noise on CM was lower for tungsten (49 HU) and ...
Source: Investigative Radiology - Category: Radiology Tags: Original Articles Source Type: research