Interlaboratory comparison investigations (ICIs) and external quality assurance schemes (EQUASs) for flame retardant analysis in biological matrices: Results from the HBM4EU project

Environ Res. 2021 Jul 20:111705. doi: 10.1016/j.envres.2021.111705. Online ahead of print.ABSTRACTThe European Human Biomonitoring Initiative (HBM4EU) is coordinating and advancing human biomonitoring (HBM). For this purpose, a network of laboratories delivering reliable analytical data on human exposure is fundamental. The analytical comparability and accuracy of laboratories analysing flame retardants (FRs) in serum and urine were investigated by a quality assurance and control (QA/QC) scheme comprising interlaboratory comparison investigations (ICIs) and external quality assurance schemes (EQUASs). This paper presents the evaluation process and discusses the results of four ICI/EQUAS rounds performed from 2018 to 2020 for the determination of ten halogenated flame retardants (HFRs) represented by three congeners of polybrominated diphenyl ethers (BDE-47, BDE-153 and BDE-209), two isomers of hexabromocyclododecane (α-HBCD and γ-HBCD), two dechloranes (anti-DP and syn-DP), tetrabromobisphenol A (TBBPA), decabromodiphenylethane (DBDPE) and 2,4,6-tribromophenol (2,4,6-TBP) in serum and four metabolites of organophosphorus flame retardants (OPFRs) in urine, at two concentration levels. The number of satisfactory results reported by laboratories increased during the four rounds. In the case of HFRs, the scope of the participating laboratories varied substantially (from two to ten) and in most cases did not cover the entire target spectrum of chemicals. The highest participatio...
Source: Environmental Research - Category: Environmental Health Authors: Source Type: research