Candidate reference measurement procedure based on HPAEC-PAD for the development of certified reference materials for monosaccharides in serum

Anal Bioanal Chem. 2021 Jul 20. doi: 10.1007/s00216-021-03537-8. Online ahead of print.ABSTRACTTo achieve the measurement reliability of monosaccharides used as diagnostic markers in clinical fields, it is essential to establish certified reference materials (CRMs). The purpose of this study is to develop a serum CRM by adopting high-performance anion-exchange chromatography with pulsed amperometric detection (HPAEC-PAD) as a new candidate reference measurement procedure for the measurement of glucose and galactose, common diagnostic markers of diabetes and galactosemia, respectively. Using various monosaccharides as internal standards, the accuracy of the HPAEC-PAD method was tested by measuring glucose CRM following treatment with three different deproteinization methods: ultrafiltration, protein precipitation by trichloroacetic acid (TCA), and protein precipitation by acetonitrile. Results showed that ultrafiltration and 5% TCA provided good accuracy with every tested monosaccharide as the internal standard. Accordingly, serum samples in this study were treated by ultrafiltration after adding 2-deoxy-D-glucose and arabinose, which were selected as internal standards for galactose and glucose, respectively. Both intra- and inter-day recovery tests showed good precision and accuracy within 2%. From the serum CRM batches prepared at two levels, 11 units were analyzed by exact-matched calibration methods, and the mass fractions of galactose and glucose were determined via HPAE...
Source: Analytical and Bioanalytical Chemistry - Category: Chemistry Authors: Source Type: research