High Peak Inspiratory Pressures After a Change of Heat and Moisture Exchangers: A Case Series

We describe 4 cases of gradually increasing PIPs from progressive HME obstruction that occurred within 6 months of an institutional change to a newer model. HME obstruction should be considered when PIPs are elevated. Alternative heat and moisture preservation strategies should be contemplated, as HMEs increase airflow resistance and add another point of possible malfunction.
Source: A&A Case Reports - Category: Anesthesiology Tags: Case Series Source Type: research