Evaluation of 4-Hz log files and secondary Monte Carlo dose calculation as patient-specific quality assurance for VMAT prostate plans

CONCLUSION: SciMoCa calculations of Monaco RT plans and RT log file plans are in excellent agreement to each other. Therefore, 4-Hz log files and SciMoCa can replace labor-intensive phantom-based measurements as patient-specific QA.PMID:34151502 | DOI:10.1002/acm2.13315