Placement of finger oximeter on the ear: comparison with oxygen saturation values taken from the finger

This study evaluated how treatment decisions might be impacted by SpO2 values obtained using a finger probe placed on the pinna of the ear in a cohort of 46 patients receiving non-invasive ventilation compared with values obtained from a probe on the finger and the results of arterial blood gas (ABG) (SaO2) analysis. Bland-Altman analysis was performed to evaluate agreement between the methods. Finger probe saturation was not statistically different from SaO2, with a mean difference of -0.66% (P>0.05). Saturation from the ear was significantly different (-4.29%; P<0.001). Subgroup analysis in hypoxic patients (SaO2<90%) showed a significant difference between ABG SaO2, and finger and ear SpO2. The study provides evidence that placement of a finger probe on the ear is unsafe clinical practice, potentially leading to patient mismanagement.PMID:34109812 | DOI:10.12968/bjon.2021.30.11.666
Source: British Journal of Nursing - Category: Nursing Authors: Source Type: research