On the MLC leaves alignment in the direction orthogonal to movement

J Appl Clin Med Phys. 2021 May 30. doi: 10.1002/acm2.13267. Online ahead of print.ABSTRACTThe main focus of the recommended spatial accuracy tests for the multi-leaf collimators (MLC) is calibration of the leaf position along the movement direction and overall alignment to the radiation isocenter. No explicit attention was typically paid to the alignment of the leaves from the opposing banks in the direction orthogonal to movement. This paper is a case study demonstrating that verification of such alignment at the time of acceptance testing is prudent. The original standard MLC (SMLC) on an MRIdian MRI-guided linac (ViewRay Inc., Mountain View, CA, USA) was upgraded to a high-speed MLC (HSMLC), which is supposed to be mechanically identical to the SMLC except for the higher drive screw pitch. The results of the end-to-end IMRT tests demonstrated unacceptable dosimetric results exemplified by an average and maximum ion chamber (IC) point dose error in the high-dose low-gradient region of 2.5 ± 1.4% and 4.6%, respectively. Before the upgrade, those values were 0.3 ± 0.7% and 0.9%, respectively. An exhaustive analysis of possible failure modes eventually zeroed in on the average misalignment of about 1 mm in the Y (along the couch) direction between the right and left upper MLC banks. The MLC was replaced, reducing the Y-direction misalignment to 0.4 mm. As a result, the average and maximum IC dose-errors became acceptable 1.0 ± 0.7% and 1.6%, respectively. Simple film and/or...