Investigation of absolute dose calibration accuracy for TomoTherapy using real water

J Appl Clin Med Phys. 2021 May 31. doi: 10.1002/acm2.13311. Online ahead of print.ABSTRACTA systematic bias in TomoTherapy output calibration was reported by the Imaging and Radiation Oncology Core Houston (IROC-H) after analyzing intensity-modulated radiation therapy (IMRT) credentialing results from hundreds of TomoTherapy units. Multiple theories were developed to explain this observation. One theory was that the use of a solid water "cheese" phantom instead of real water in the calibration measurement was the culprit. A phantom filled with distilled water was built to investigate whether our TomoTherapy was miscalibrated due to the use of a solid water phantom. A miscalibration of -1.47% was detected on our TomoTherapy unit. It is found that despite following the vendor's updated recommendation on computed tomography (CT) number to density calibration, the cheese phantom was still mapped to a density of 1.028 g/cm3 , rather than the 1.01 g/cm3 value reported in literature. When the density of the cheese phantom was modified to 1.01 g/cm3 in the treatment planning system, the measurement also indicated that our TomoTherapy machine was miscalibrated by -1.52%, agreeing with the real water phantom findings. Our single-institution finding showed that the cheese phantom density assignment can introduce greater than 1% errors in the TomoTherapy absolute dose calibration. It is recommended that the absolute dose calibration for TomoTherapy be performed either in real water or in...