Correlation of iCare ic100 tonometry with iCare TA01i in screening of unselected population in Northern Finland Birth Cohort Eye study

CONCLUSION: In summary, we found ic100 rebound tonometry to be both reliable and effective, although CCT may influence IOP measurements with ic100 and TA01i. Therefore, iCare ic100 is suitable for IOP measurement in large cohort studies.PMID:33951959 | DOI:10.1177/11206721211014387
Source: European Journal of Ophthalmology - Category: Opthalmology Authors: Source Type: research