Reducing fire and burn risk in the operating room —testing of a novel device

ConclusionsThe Gloshield device is effective in reducing the risk of thermal burn injury by protecting the distal ends of endoscopic light cables from operating room materials. However, the device needs to be attached appropriately in order to provide protective benefits.
Source: Surgical Endoscopy - Category: Surgery Source Type: research
More News: Burns | Endoscopy | Study | Surgery