Can Finger-Generated Force Be Used Reliably to Connect the Transducer for Resonance Frequency Analysis in Determining Implant Stability?

CONCLUSION: For an accurate measurement of ISQ, the minimum torque necessary to insert the transducer into the implant for Osstell was 10 Ncm, while for Penguin, it was 5 Ncm. Therefore, when using Osstell to assess implant stability, the authors suggest the use of a torque wrench to ensure 10 Ncm of force is applied when tightening the transducer into the implant to obtain accurate stability measurements. When using Penguin, the maximum finger-generated tightening force is enough. PMID: 33270054 [PubMed - in process]
Source: International Journal of Oral and Maxillofacial Implants - Category: ENT & OMF Tags: Int J Oral Maxillofac Implants Source Type: research