Atrial fibrillation detection by the subcutaneous defibrillator: real-world clinical performances and implications from a multicentre study

ConclusionThe S-ICD AF detection algorithm yields a high sensitivity for AF diagnosis. Low specificity and positive predictive value contribute to a high remote monitoring-notification workload and underline the necessity of a manual analysis. Atrial fibrillation diagnosis by the S-ICD AF detection algorithm might lead to significant therapeutic adjustments.
Source: Europace - Category: Cardiology Source Type: research