Sensors, Vol. 20, Pages 6128: An Integrated Front-end Circuit Board for Air-Coupled CMUT Burst-Echo Imaging

Sensors, Vol. 20, Pages 6128: An Integrated Front-end Circuit Board for Air-Coupled CMUT Burst-Echo Imaging Sensors doi: 10.3390/s20216128 Authors: Lei Ye Jian Li Hui Zhang Dongmei Liang Zhuochen Wang To conduct burst-echo imaging with air-coupled capacitive micromachined ultrasonic transducers (CMUTs) using the same elements in transmission and reception, this work proposes a dedicated and integrated front-end circuit board design to build an imaging system. To the best of the authors’ knowledge, this is the first air-coupled CMUT burst-echo imaging using the same elements in transmission and reception. The reported front-end circuit board, controlled by field programmable gate array (FPGA), consisted of four parts: an on-board pulser, a bias-tee, a T/R switch and an amplifier. Working with our 217 kHz 16-element air-coupled CMUT array under 100 V DC bias, the front-end circuit board and imaging system could achieve 22.94 dB signal-to-noise ratio (SNR) in burst-echo imaging in air, which could represent the surface morphology and the three-dimensional form factor of the target. In addition, the burst-echo imaging range of our air-coupled CMUT imaging system, which could work between 52 and 273 mm, was discussed. This work suggests good potential for ultrasound imaging and gesture recognition applications.
Source: Sensors - Category: Biotechnology Authors: Tags: Letter Source Type: research