Development and accuracy evaluation of a degree of occlusion visualization system for roller pumps used in cardiopulmonary bypass

In this study, we have developed a degree of occlusion quantification system which constructs the flow channel shape within an occluded tube from red optical density images, and we have verified the validity of this system. Utilizing a linear actuator, an acrylic roller and raceway, a solution colored with simulated blood powder, and a 3/8-inch vinyl chloride tube, this system uses a camera to capture red optical density images within an occluded tube and constructs the tube flow channel shape using a formula manipulation system. To verify the accuracy of this system, we compared the thickness of a cross-section of the flow channel constructed with the degree of occlusion quantification system with the thickness of a cross-section of silicone cured under the same occlusion conditions. Our experiments indicated that for areas with a small tube gap, this system can construct highly accurate three-dimensional shapes and obtain quantitative indicators assessing the degree of occlusion.
Source: Journal of Artificial Organs - Category: Transplant Surgery Source Type: research