A Prospective Analysis of iStent Inject Microstent Implantation: Surgical Outcomes, Endothelial Cell Density, and Device Position at 12 Months

Precis: The surgical outcome of iStent inject devices is associated with device protrusion within the anterior chamber. Schlemm canal (SC) dilatation has a prognostic value. iStent inject devices do not move within the first year after implantation. Introduction: The iStent inject is a device designed to be implanted ab-interno through the trabecular meshwork. The present study follows up on our preliminary report, using successive in vivo anterior segment optical coherence tomography (AS-OCT) imaging to analyze the associations between stent positioning, iridocorneal angle structures, and surgical outcomes. Methods: In total, 54 eyes of 42 patients (73.3±7.4 y) with cataract and mild-to-moderate open-angle glaucoma were examined in this prospective study. All patients underwent implantation of 2 iStent inject devices combined with phacoemulsification. Patients were followed up over a 12-month period. AS-OCT was performed after 3 and 12 months. Thirty unoperated fellow eyes served as control eyes. Results: Intraocular pressure (IOP) decreased from 16.5±4.2 mm Hg at baseline to 15.1±3.7 mm Hg at 12 months (−8.7%; P=0.004), while medications decreased from 1.8±1.0 to 0.5±0.9 (−72.2%; P