Multi-energy CT imaging for large patients using dual-source photon-counting detector CT.

Multi-energy CT imaging for large patients using dual-source photon-counting detector CT. Phys Med Biol. 2020 Jun 05;: Authors: Tao S, Marsh JF, Tao A, Michalak GJ, Rajendran K, McCollough CH, Leng S Abstract Multi-energy CT imaging of large patients with conventional dual-energy (DE)-CT using an energy-integrating-detector (EID) is challenging due to photon starvation-induced image artifacts, especially in lower tube potential (80-100 kV) images. Here, we performed phantom experiments to investigate the performance of DECT for morbidly obese patients, using an iodine and water material decomposition task as an example, on an emulated dual-source (DS)-photon-counting-detector (PCD)-CT, and compared its performance with a clinical DS-EID-CT. An abdominal CT phantom with iodine inserts of different concentrations was wrapped with tissue-equivalent gel layers to emulate a large patient (50 cm lateral size). The phantom was scanned on a research whole-body single-source (SS)-PCD-CT (140 kV tube potential), a DS-PCD-CT (100/Sn140 kV; Sn140 indicates 140 kV with Sn filter), and a clinical DS-EID-CT (100/Sn140 kV) with the same radiation dose. Phantom scans were repeated five times on each system. The DS-PCD-CT acquisition was emulated by scanning twice on the SS-PCD-CT using different tube potentials. The multi-energy CT images acquired on each system were then reconstructed, and iodine- and water-specific images were generated using mater...