Single Angio-Seal vascular closure device for transfemoral access exceeding 8F.

CONCLUSIONS: In selected cases, without availability of appropriate VCDs, a single Angio-SealTM-VIP 8F served as a feasible option to achieve sufficient hemostasis of transfemoral access exceeding 8F without relevant bleeding complications. Very low BMI, tall body-height, and severe atherosclerosis were identified to predispose to access site complications. PMID: 32402666 [PubMed - as supplied by publisher]
Source: Journal of Cardiology - Category: Cardiology Authors: Tags: J Cardiol Source Type: research