Image quality of list-mode proton imaging without front trackers.

Image quality of list-mode proton imaging without front trackers. Phys Med Biol. 2020 Apr 28;: Authors: Sølie JRR, Volz L, Pettersen HESES, Piersimoni P, Odland OH, Roehrich D, Helstrup H, Peitzmann T, Ullaland K, Varga-Kofarago M, Mehendale S, Grøttvik OS, Eikeland VN, Meric I, Seco J Abstract List mode proton imaging relies on accurate reconstruction of the proton most likely path (MLP) through the patient. This typically requires two sets of position sensitive detector systems, one upstream (front) and one downstream (rear) of the patient. However, for a clinical implementation it can be preferable to omit the front trackers (single-sided proton imaging). For such a system, the MLP can be computed from information available through the beam delivery system and the remaining rear tracker set. In this work, we use Monte Carlo simulations to compare a conventional double-sided (using both front and rear detector systems) with a single-sided system (only rear detector system) by evaluating the spatial resolution of proton radiographs (pRad) and proton CT images (pCT) acquired with these set-ups. Both the pencil beam spot size, as well as the spacing between spots was also adjusted to identify the impact of these beam parameters on the image quality. Relying only on the pencil beam central position for computing the MLP resulted in severe image artifacts both in pRad and pCT. Using the recently extended-MLP formalism that incorporate...