The effect of megavoltage field size on intrafraction cone-beam CT image quality

AbstractTo investigate the effects of scatter from a megavoltage treatment beam on intrafraction cone beam CT (CBCT) image quality. The effects of treatment beam field size and phantom geometry were investigated as well as the clinical success of IFI. Intrafraction imaging (IFI) was performed on four phantoms with four different MV field sizes using a 6 MV FFF source. The image quality of the intrafraction CBCT images was compared to that of a baseline CBCT (i.e. with no treatment beam on) and quantified using noise and low contrast visibility. Increasing the kV tube current was explored as a possible method to reduce noise induced by the MV photon scatter in the intrafraction-CBCTs. The clinical success of all IFI patients over a 2 month period was reviewed. Intrafraction-CBCT image quality and low-contrast visibility deteriorated as MV field size increased. The extent of image degradation was found to depend on the mass of the phantom resulting in a more pronounced effect for a pelvic phantom than a thoracic phantom. While increasing the tube current could reduce the noise in the intrafraction-CBCT images, increasing the current by a factor of 4 failed to reach baseline image quality. Anatomy was found to be the primary indication of clinical IFI failure with all observed failures occurring during abdominal treatments. Image quality was found to decrease with increasing MV field size and decrease with increasing treatment anatomy mass. When considering intrafraction imaging...