Automated Ultrasound Measurement of the Inferior Vena Cava: An Animal Study.

In this study, we developed a cross-correlation (CC)-based method for automated IVC movement tracking and measurement, which requires minimal manual input and avoids manual selection of reference markers. Based on the CC method, two approaches, named direct and relative approaches, were used to calculate the maximum, minimum, and variation of the IVC diameter, and compared with the manual measurement. Fifty-four ultrasound cine-loops collected from nine pigs were tested. The results reveal that both the proposed approaches were well agreed with the manual measurement. The errors of the direct approach were less than 9%, while those of relative approach were as high as 26.7%. It is concluded that the proposed direct approach is superior for IVC diameter measurements, which can be comparable with manual counterpart, serving as an alternative to traditional IVC measurement. PMID: 32238107 [PubMed - as supplied by publisher]