Measurement of Physical Parameters and Development of a Light Emitting Diodes Device for Therapeutic Use

The objective of this study was to develop a light-emitting device using LED, with proper measurements for application in clinical research. It was used 267 LEDs, powered with 12-V voltage and fixed on a plate of ethylene-vinyl acetate (25  × 42 cm), equidistant at 1.0 cm. For the calculation of red and infrared irradiation, a spectrometer was used, and the data were processed in routines implemented in the OriginPro 8.5.0 SR1 Software. The irradiance was determined by the integration of the spectral irradiation in the LED emiss ion region. The red LED has a wavelength of 620 ± 10 nm, a power density of 52.86 mW/cm2, power of 6.6  mW, and total power of 1.76 W on the device. The infrared LED has a wavelength of 940 ± 10 nm, power density 33.7 mW/cm2, power of 6  mW, and total power of 1.6 W on the device. The LED characterization enables the generation and application of energy with greater precision and reproducibility. Besides, it is a light source, a device capable of framing large areas, reducing the time and cost of the application in different clin ical conditions related to neuromuscular performance or rehabilitation.