Online monitoring of end-tidal propofol in balanced anesthesia by anisole assisted positive photoionization ion mobility spectrometer.

Online monitoring of end-tidal propofol in balanced anesthesia by anisole assisted positive photoionization ion mobility spectrometer. Talanta. 2020 May 01;211:120712 Authors: Jiang D, Chen C, Wang X, Li M, Xiao Y, Liu Y, Li E, Li H Abstract Online measuring end-tidal propofol concentration during balanced anesthesia is important for anesthetists to learn the patient's anesthesia depth as exhaled propofol concentration is well related to blood propofol concentration. In previous work, exhaled propofol was detected using acetone assisted negative photoionization ion mobility spectrometer, however, the existence of high concentration sevoflurane interfered the response of propofol. In this work, an anisole assisted photoionization ion mobility spectrometer operated in positive mode was developed to sensitively and selectively measure the end-tidal propofol by eliminating the interferences of exhaled humidity and sevoflurane during balanced anesthesia. Anisole molecular ion is stable enough not to go under proton transfer reaction with water presents in the exhaled breath. Hence, the exhaled humidity related peaks were eliminated and only one propofol product ion peak (K0 = 1.50 cm2 V-1 s-1) was observed. The relative standard deviation (RSD) ranging from 0.64%-0.91% showed good repeatability and the quantitative range was 0.2-40 ppbv with a response time of 4 s. Finally, the performance of the proposed method was demonstrated by...
Source: Talanta - Category: Chemistry Authors: Tags: Talanta Source Type: research