2019 HRS/EHRA/APHRS/LAHRS focused update to 2015 expert consensus statement on optimal implantable cardioverter-defibrillator programming and testing

AbstractThe2015 HRS/EHRA/APHRS/SOLAECE Expert Consensus Statement on Optimal Implantable Cardioverter-Defibrillator Programming and Testing provided guidance on bradycardia programming, tachycardia detection, tachycardia therapy, and defibrillation testing for implantable cardioverter-defibrillator (ICD) patient treatment. The 32 recommendations represented the consensus opinion of the writing group, graded by Class of Recommendation and Level of Evidence. In addition, Appendix B provided manufacturer-specific translations of these recommendations into clinical practice consistent with the recommendations within the parent document. In some instances, programming guided by quality evidence gained from studies performed in devices from some manufacturers was translated such that this programming was approximated in another manufacturer ’s ICD programming settings. The authors found that the data, although not formally tested, were strong, consistent, and generalizable beyond the specific manufacturer and model of ICD. As expected, because these recommendations represented strategic choices to balance risks, there have been repor ts in which adverse outcomes were documented with ICDs programmed to Appendix B recommendations. The recommendations have been reviewed and updated to minimize such adverse events. Notably, patients who do not receive unnecessary ICD therapy are not aware of being spared potential harm, whereas pati ents in whom their ICD failed to treat life-threate...