A Low-Cost End-to-End sEMG-Based Gait Sub-Phase Recognition System

This study presents a low-cost but effective end-to-end sEMG-based gait sub-phase recognition system, which contains a wireless multi-channel signal acquisition device simultaneously collecting sEMG of thigh muscles and plantar pressure signals, and a novel neural network-based sEMG signal classifier combining long-short term memory (LSTM) with multilayer perceptron (MLP). We evaluated the system with subjects walking under five conditions: flat terrain at 5 km/h, flat terrain at 3 km/h, 20 kg backpack at 5 km/h, 20 kg shoulder bag at 5 km/h and 15° slope at 5 km/h. Experimental results show that the proposed method achieved average classification accuracies of 94.10%, 87.25%, 90.71%, 94.02%, and 87.87%, respectively, which were significantly higher than existing recognition methods. Additionally, the proposed system had a good real-time performance with low average inference time in the range of 3.25 ~ 3.31 ms.
Source: IEE Transactions on Neural Systems and Rehabilitation Engineering - Category: Neuroscience Source Type: research