Quantification and homogenization of image noise between two CT scanner models.

Quantification and homogenization of image noise between two CT scanner models. J Appl Clin Med Phys. 2019 Dec 20;: Authors: Einstein SA, Rong XJ, Jensen CT, Liu X Abstract Feedback from radiologists indicated that differences in image appearance and noise impeded reading of post-contrast computed tomography (CT) scans from an updated CT scanner that was recently added to a fleet of existing scanners from the same vendor, despite using identically named reconstruction algorithms. The goals of this work were to quantify and possibly standardize image quality on the new and an existing scanner using phantom images. Three months of daily quality control images were analyzed to determine the mean CT number and noise magnitude in a water phantom. Next, subtraction images from the uniformity section of an American College of Radiology CT phantom were used to generate noise power spectra for both scanners. Then, a semi-anthropomorphic liver phantom was imaged with both scanners in triplicate using identical body protocols to quantify differences CT number and noise magnitude. Finally, the scanner dependence of CT number and noise magnitude on material attenuation was quantified using a multi-energy CT phantom with 15 material inserts. Significant differences between scanners were determined using a paired or Welch's t test as appropriate. In daily quality control images, the new scanner exhibited slightly higher CT number (0.697 vs. 0.412, ...
Source: Journal of Applied Clinical Medical Physics - Category: Physics Authors: Tags: J Appl Clin Med Phys Source Type: research