Optimisation of the drill-in behaviour of the EcoFit ® SC threaded cup.

Optimisation of the drill-in behaviour of the EcoFit® SC threaded cup. Biomed Tech (Berl). 2019 Dec 13;: Authors: Thorwächter C, Woiczinski M, Santos I, Schmidutz F, Paulus A, Jansson V, Weber P Abstract Threaded cups show good clinical results when implanted correctly. In clinical use, multiple cases with an incomplete placement of the EcoFit threaded cup (implantcast) were observed. This behaviour could not be explained intra- and postoperatively. The aim of this study was to compare and optimise the drill-in-behaviour of the EcoFit cup in a biomechanical investigation. EcoFit cup sizes 46, 50 and 54 mm were compared with the SC cup (Aesculap) size 50 mm. Foam blocks (Sawbones) of density 0.16 g/ml (pcf 10), 0.32 g/ml (pcf 20) and 0.48 g/ml (pcf 30) were used. After standardised placement using a universal testing system (n = 8 per group), the primary stability, the overhang of the cups and the drill-in behaviour were measured. Overreamings of 1 and 2 mm were performed (pcf 20, n = 8) for the EcoFit cup size 50 and the primary stability as well as the overhang measurements were examined. Measurements of the cup diameter, thread depth and thread pitch were performed on three-dimensional (3D) images of the cup size 50 mm. The drill-in behaviour was different between the EcoFit and the SC cups. Even with maximum torque, the EcoFit cup could not be positioned as deep as the SC cup in standard reaming conditions (overhang of 1...
Source: Biomedizinische Technik/Biomedical Engineering - Category: Biomedical Engineering Tags: Biomed Tech (Berl) Source Type: research