Converting Coherence to Signal-to-noise Ratio for Enhancement of Adaptive Ultrasound Imaging.

In this study, a new method based on the signal-to-noise ratio (SNR) among the element echo signals was developed for enhancement of the effect of the coherence factor (CF), which was previously developed for improvement in spatial resolution and contrast. In the proposed method, a new factor, namely, SNR factor, was introduced, and the relationship between the previously developed CF and SNR factor was discussed. The proposed method was implemented in plane wave imaging, and the performance was evaluated by simulated and phantom experiments. In simulation, the lateral spatial resolution and contrast obtained with the conventional CF were 0.23 mm and 47.0 dB, respectively, which were significantly better than 0.39 mm and 15.3 dB obtained by conventional delay-and-sum (DAS) beamforming. Using the proposed method, the lateral spatial resolution and contrast were further improved to 0.12 mm and 69.8 dB, respectively. Similar trends were found also in phantom experiments. PMID: 31802696 [PubMed - as supplied by publisher]
Source: Ultrasonic Imaging - Category: Radiology Tags: Ultrason Imaging Source Type: research
More News: Radiology | Study | Ultrasound