Functional and radiographic evaluation of an interspinous device as an adjunct for lumbar interbody fusion procedures.

Functional and radiographic evaluation of an interspinous device as an adjunct for lumbar interbody fusion procedures. Biomed Tech (Berl). 2019 Oct 24;: Authors: Spicher A, Schmoelz W, Schmid R, Stofferin H, Craig NJA Abstract In the last decades, several interspinous process devices were designed as a minimally invasive treatment option for spinal stenosis. In order to minimise surgical trauma, interspinous process devices were recently discussed as an alternative posterior fixation in vertebral interbody fusions. Therefore, the purpose of this study was to evaluate the effect of a newly designed interspinous device with polyester bands (PBs) on range of motion (RoM) and centre of rotation (CoR) of a treated motion segment in comparison with an established interspinous device with spikes (SC) as well as with pedicle screw instrumentation in lumbar fusion procedures. Flexibility tests with an applied pure moment load of 7.5 Nm were performed in six monosegmental thoracolumbar functional spinal units (FSUs) in the following states: (a) native, (b) native with PB device, (c) intervertebral cage with PB device, (d) cage with SC and (e) cage with internal fixator. The resulting RoM was normalised to the native RoM. The CoR was determined of X-ray images taken in maximal flexion and extension during testing. In flexion and extension, the PB device without and with the cage reduced the RoM of the native state to 58% [standard deviation (SD...
Source: Biomedizinische Technik/Biomedical Engineering - Category: Biomedical Engineering Tags: Biomed Tech (Berl) Source Type: research