Novel Application Software for the Semi-Automated Analysis of Infrared Meibography Images

Purpose: To develop semi-automated application software that quickly analyzes infrared meibography images taken with the CSO Sirius Topographer (CSO, Italy) and to compare them to the manual analysis system on the device (Phoenix software platform). Methods: A total of 52 meibography images verified as high quality were used and analyzed through manual and semi-automated meibomian gland (MG) detector software in this study. For the manual method, an experienced researcher circumscribed the MGs by putting dots around grape-like clusters in a predetermined rectangular area, and Phoenix software measured the MG loss area by percentage, which took around 10 to 15 minutes. MG loss was graded from 1 (75%). For the semi-automated method, 2 blind physicians (I and II) determined the area to be masked by putting 5 to 6 dots on the raw images and measured the MG loss area using the newly developed semi-automated MG detector application software in less than 1 minute. Semi-automated measurements were repeated 3 times on different days, and the results were evaluated using paired-sample t test, Bland–Altman, and kappa κ analysis. Results: The mean MG loss area was 37.24% with the manual analysis and 40.09%, 37.89%, and 40.08% in the first, second, and third runs with the semi-automated analysis (P