[Problems and chances of high field magnetic resonance imaging].

[Problems and chances of high field magnetic resonance imaging]. Radiologe. 2013 May;53(5):401-10 Authors: Ladd ME, Bock M Abstract CLINICAL/METHODICAL ISSUE: The spatial, temporal and spectral resolution in magnetic resonance imaging (MRI) is in many cases currently not sufficient to detect submillimeter lesions or to image the dynamics of the beating heart. STANDARD RADIOLOGICAL METHODS: At present MRI systems at 1.5 T and 3 T are the standard units for clinical imaging. METHODICAL INNOVATIONS: The use of ultrahigh magnetic fields of 7 T and higher increases the signal-to-noise ratio, which holds promise for a significant improvement of the spatial and/or temporal resolution as well as for new contrast mechanisms. PERFORMANCE: With 7 T MRI, images of the brain have been acquired routinely with a spatial resolution of 0.3 mm. The theoretical improvement of the signal-to-noise ratio is often not fully realized due to B1 inhomogeneities and contrast variations. ACHIEVEMENTS: With MRI at 7 T a notable increase in spatial resolution can be achieved. Methods such as time-of-flight MR angiography and susceptibility-weighted imaging (e.g. neurofunctional MRI, fMRI) profit especially from the higher field strengths. Transmission field inhomogeneities are still a major challenge for ultrahigh field (UHF) MRI and are also a partially unsolved safety problem. PRACTICAL RECOMMENDATIONS: The use of UHF MRI is currently limited to special application...
Source: Der Radiologe - Category: Radiology Authors: Tags: Radiologe Source Type: research