Error detection using an electromagnetic tracking system in multi-catheter breast interstitial brachytherapy.

Error detection using an electromagnetic tracking system in multi-catheter breast interstitial brachytherapy. Phys Med Biol. 2019 Sep 10;: Authors: Masitho S, Kallis K, Strnad V, Fietkau R, Bert C Abstract The hybrid treatment delivery system (HTDS) has been proposed as a possible option for a quality assurance in the multi-catheter interstitial brachytherapy for breast cancer patients. The system, which consists out of a prototype afterloader with an integrated electromagnetic tracking (EMT) sensor and an EMT system, allows the automatic measurement of implanted catheters.
 
 To test the feasibility of the system for error detection, possible treatment planning errors and treatment delivery errors were simulated. Planning errors such as an incorrect offset value, an incorrect indexer length, tip/connector end swaps, and partial swaps, and; treatment delivery errors such as catheter shifts and catheter connection swaps were manually simulated using phantoms. An in-house Matlab routine was used to assess geometrical deviations between the dwell positions defined based on CT and EMT measurement. Additionally, the influence of implant motion on the detection ability of the system was assessed. An algorithm for the detection and specification of errors based on the error simulation results was developed. At the University Hospital Erlangen, a patient study is ongoing, where errors in patient data were analyzed usi...
Source: Physics in Medicine and Biology - Category: Physics Authors: Tags: Phys Med Biol Source Type: research